Test for determining Silt Density Index
Equipment |
- Millipore® Filter Pad Holder with a 0.45 micron filter disc.
- 500 ml. Graduated cylinder
- Stopwatch
- Dull Tweezers
Procedure |
- Unbolt the filter holder, wet it, and place a 0.45 micron filter on the back-up plate using the dull tweezers.
- Place the "O"-ring properly and then replace the top half of the filter holder and bolt loosely.
- Connect apparatus to the feedwater line (minimum 30 PSIG) in a vertical, downflow position and open the needle valve a crack.
- Loosen two adjacent filter-holder bolts, tilt the apparatus and bleed out all trapped air, then retighten bolts and adjust the pressure to 30 PSI while starting the stopwatch.
- Immediately run the flow into the graduated cylinder and measure the time required to collect 500 ml, maintaining a pressure of 30 PSIG during the run. Mark the time (t1) and keep the watch running.
- Repeat the previous step immediately after 5, 10, and/or 15 minutes of total elapsed time. Mark the collection times (t5, t10, tF).
- Close the sample valve and disconnect he apparatus after the 15 minute reading. Remove the disk and place it in a plastic bag for later examination.
Calculation |
- Calculate the ratios R = t1/t5 , t1/t10 and/or t1/tF.
- Estimate % P30 at the selected time(s) from the following equation:
% P30 = [1-R] × 100. - Estimate the Silt Density Index, SDI, from the following equation:
- SDI = % P30÷ Elapsed Time (minutes)